Roche/Hitachi cobas c 311 analyzer Specification of interface
Host Communication Setting
Roche Diagnostics May 2011
Host Interface Manual – Version 1.3 B-7
Parameter Option
Ignore host orders for samples
with already existing 1st results
check box On/Off If this funciton is enable, when orders are sent from Host,
orders for tests with already existing 1st run results are ignored
and not measured again.
Additional orders are detected and processed and Automatic
rerun triggered by system instrument is possible.
Rerun and 3rd result orders from host are not possible.
If this function is disable, all orders are detected and measured
even if not necessary.
Note: For example, during S.B.S. mode, the inquiry of orders
to Host is performed repeatedly.When this function is
enable, orders sent from Host which have 1st result are not
measured.
System
Name used on reports and in communications for the
cobas c 311 analyzer.
Host
Name used on reports and in communications for the host
system.
System ID
The ID number used to identify the system in communications
with the host.
Host ID
The ID number used to identify the host in communications
with the analyzer.
Text Setting
Result Only check box On/Off This function applies to all Realtime Communications. This
does not apply to the Batch Mode. When this function is
enabled, communication during analysis is limited only to the
analytical data transmissions. Inquiries for the Test Selection
Information or Auto Rerun Selection Information are not made.
Test selections would need to be made either manually by the
operator at the analyzer, by Batch Mode from the host, or by
use of the Default Profile.
TS Timeout check box On/Off
and input 10-9999
This function allows the operator to set a timeout interval for
Test Selection Information Inquiry for Realtime
Communications and for routine, Stat and rerun samples. The
specified timeout interval determines how long the analyzer will
wait for a ‘Response from the host’. The range is 10-9999
seconds (depending on TS Priority Mode), with a default of 18
seconds. Enabling this function optimizes the throughput of the
system. If the function is not enabled, the analyzer will wait
indefinitely for Test Selection Information.
Auto Rerun TS check box On/Off This function applies to Auto Rerun TS inquiry within the
Realtime Communications if the upload setting is not “by
sample. When this function is enabled in the Auto Rerun mode
during analysis, an inquiry is made for the Auto Rerun
selections. When this function is not enabled, no inquiry for the
Auto Rerun Selection is made.
Auto Rerun TS Timeout check box On/Off
and input 10-9999
This function applies to the reception of Auto Rerun Test
Selection Information in Realtime Communications. The range
is 10-9999 seconds, with a default of 18 seconds (depending
on TS Priority Mode). If the function is not enabled, the
analyzer waits indefinitely for Rerun Test Selection
Information.
Manual Rerun T S check box On/Off This function applies to the Test Selection Information inquiry
for Manual Rerun samples in Realtime Communication. When
this function is enabled, an inquiry is made for the test
selections for rerun samples during analysis. Any samples that
have previously been processed create a TS inquiry to the
host. When this function is not enabled, no TS inquiry for
Manual Rerun samples is made.
Stat TS check box On/Off This function applies to the Test Selection Information inquiry
for Stat samples in Realtime Communications. When this
function is enabled, an inquiry is made for the test selections
for Stat samples during analysis. When this function is not
enabled, no TS inquiry for Stat samples is made.