Roche/Hitachi cobas c 311 analyzer Specification of interface
Host Communication Setting
Roche Diagnostics May 2011
Host Interface Manual – Version 1.3 B-8
Parameter Option
TS Inquire Always check box On/Off This function applies to the Test Selection Information inquiry
for routine and Stat samples in Realtime Communications.
This does not apply to TS information inquiry for rerun
samples, Auto Rerun and Manual Rerun. When this function is
enabled, an inquiry is made whether the TS information is in
the analyzer or not. When not enabled, an inquiry is made only
for the samples that do not have test selections.
TS Ask in Barcode Read
Error
check box On/Off If this function is enabled, and the analyzer cannot read a
sample barcode, “***...” [13 or 22 characters of ‘*’] is sent as
the sample ID, and the rack ID and position are sent to the
host as key information to identify the sample. If the host has
the table of the rack ID, position and Sample ID, the host can
send back to the analyzer the correct Sample ID instead of
‘***...’
Note
Because Host protocol and text field is same as cobas6000,
Rack ID field is used in cobas c 311 analyzer communication
text.
The only currently known use for this function is with a Sample
Handling System that ‘manufactures’ sample aliquots for the
analyzer. This type of system would track samples by position
on S.Disk. Unless the Sample Handling System can perform
this function, it is strongly recommended that it not be used
This function is not visible, if the Sample Barcode Reader is
switched off.
TS Priority Mode check box On/Off
If this function is enabled, after a Test Selection inquiry the
analyzer will stop sending any information, i.e. result data, until
the analyzer receives the answer to the inquiry from the host or
the TS timeout period passes. This option applies to both the
inquiry at the bar code reader and the implied inquiry after
Result upload in “By Sample” setting.
The range of the TS Timeout and Auto Rerun TS Timeout is
from 1 to 18 seconds with this mode on. With this mode off the
TS Timeout options remain at 10 to 9999 seconds. If this mode
is enabled and the TS Timeout checkbox is off, TS timeout is
handled as 18 seconds.
Sample ID 22 Digits check box On/Off
The maximum length of the sample ID is increased from 13 to
max. 22 characters.
New Mode check box On/Off The host interface protocol description in this manual is based
on this 'New Mode' setting enabled.
If disabled the protocol is identical with the protocol of the CC
Modular software for purpose of 100% compatibility with
already connected Analyzer systems.
Concerning detail of the CC Modular host interface
specification refer to
Roche/Hitachi MODULAR System Host Interface Document
Version 1.0 Ident. No. 011632701 (US ID)
Send 1st/Rerun Information check box On/Off With this mode enabled, additional information that shows
whether a query is 1
st
or rerun measurement is added to
Request Information Record. This mode is applied to only
“Inquiry for the Requested Tests” message from the
analyzer to the host, but not applied to “Inquiry of Result”
from the host to the analyzer.
If the function is not enabled, no information is added to
Request Information Record.
Send Comments check box On/Off This function applies to host communications. When this
function is enabled, the host can send patient comments to the
analyzer. Patient comments stored in the analyzer can also be
transmitted to the host along with analytical data and Reaction
Monitor data. When this function is not enabled, no patient
comments are accepted or sent.
Send Concentration Of
Chemistry Qualitative Test
. check box On/Off This function is available for Result transfer and Reaction
Monitor transfer.
When this function is selected, measurement results are also
send to Host with Qualitative result in photometric test result
record.