EasyManuals Logo
Home>Roche>Measuring Instruments>Cobas C 311

Roche Cobas C 311 User Manual

Roche Cobas C 311
62 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #59 background imageLoading...
Page #59 background image
Roche/Hitachi cobas c 311 analyzer Appendix A - Instrument Alarms
Instrument Alarm List
Roche Diagnostics May 2011
Host Interface Manual – Version 1.3 G-7
Alarm Type Alarm No
Alarm Message Alarm Description Alarm Remedy
340 24 Host Receiving Text Error Request Information Status Code is not
effective value.
Please check Request Information Status.
340 25 Host Receiving Text Error Specimen ID is not effective value. Please check Specimen ID.
340 26 Host Receiving Text Error Sample No. is not effective value. Please check Sample No.
340 28 Host Receiving Text Error Sample Type is not effective value. Please check Sample Type.
340 30 Host Receiving Text Error Rack Position No. is not effective value. Please check Rack Position No.
340 31 Host Receiving Text Error A component pause of Universal Test ID is
insufficient.
Please check Universal Test ID.
340 32 Host Receiving Text Error Action Code & Value is not effective value. Please check Action Code & Value.
340 33 Host Receiving Text Error A component pause of User Field No. 2 is
insufficient.
Please check User Field No. 2.
340 34 Host Receiving Text Error Instrument Specimen ID is not effective value. Please check Instrument Specimen ID.
340 35 Host Receiving Text Error A component pause of Comment or Special
Instructions.
Please check Comment or Special Instructions.
340 36 Host Receiving Text Error Comment or Special Instructions is not effective
value.
Please check Comment or Special Instructions.
340 37 Host Receiving Text Error A field pause of H record is insufficient. Please check H record.
340 38 Host Receiving Text Error Comment or Special Instructions is not effective
value.
Please check Comment or Special Instructions.
340 39 Host Receiving Text Error There is no record end of H record. Please check whether '<CR>' (0x0d) is in the terminus of H record.
340 40 Host Receiving Text Error A field pause of P record is insufficient. Please check P record.
340 41 Host Receiving Text Error Sequence Number of P record is not effective
value.
Please check Sequence Number of P record.
340 44 Host Receiving Text Error There is no record end of P record. Please check whether '<CR>' (0x0d) is in the terminus of P record.
340 45 Host Receiving Text Error Application Code is not effective value. Please check Application Code.
340 47 Host Receiving Text Error A field pause of O record is insufficient. Please check O record.
340 48 Host Receiving Text Error Sequence Number of O record is not effective
value.
Please check Sequence Number of O record.
340 50 Host Receiving Text Error Instrument Specimen ID is not effective value. Please check Instrument Specimen ID.
340 57 Host Receiving Text Error There is no record end of O record. Please check whether '<CR>' (0x0d) is in the terminus of O record.
340 59 Host Receiving Text Error Sample ID is not effective value. Please check Sample ID.
340 60 Host Receiving Text Error Sample No. is not effective value. Please check Sample No.
340 62 Host Receiving Text Error Rack Position is not effective value. Please check Rack Position.
340 63 Host Receiving Text Error A field pause of Q record is insufficient. Please check Q record.
340 64 Host Receiving Text Error Sequence Number of Q record is not effective
value.
Please check Sequence Number of Q record.
340 65 Host Receiving Text Error Specimen ID is not effective value. Please check Specimen ID.
340 67 Host Receiving Text Error There is no record end of Q record. Please check whether '<CR>' (0x0d) is in the terminus of Q record.
340 68 Host Receiving Text Error A field pause of C record is insufficient. Please check C record.
340 69 Host Receiving Text Error Sequence Number of C record is not effective
value.
Please check Sequence Number of C record.
340 73 Host Receiving Text Error There is no record end of C record. Please check whether '<CR>' (0x0d) is in the terminus of C record.
340 101 Host Receiving Text Error Sample No. is outside the range. Please check Sample No.
340 102 Host Receiving Text Error Class is outside the range. Please check Class.
340 104 Host Receiving Text Error Cup Position is outside the range. Please check Cup Position.
340 105 Host Receiving Text Error First run/rerun is not specified. Please confirm whether to be the First run or rerun.