Roche/Hitachi cobas c 311 analyzer Appendix A - Instrument Alarms
Instrument Alarm List
Roche Diagnostics May 2011
Host Interface Manual – Version 1.3 G-8
Alarm Type Alarm No
Alarm Message Alarm Description Alarm Remedy
340 106 Host Receiving Text Error Application Code is outside the range. Please check Application Code.
340 107 Host Receiving Text Error Sample Volume is outside the range. Please check Sample Volume.
340 108 Host Receiving Text Error Draw Time (Year) is outside the range. Please check Draw Time (Year).
340 109 Host Receiving Text Error Draw Time (Month) is outside the range. Please check Draw Time (Month).
340 110 Host Receiving Text Error Draw Time (Day) is outside the range. Please check Draw Time (Day).
340 111 Host Receiving Text Error Draw Time (Hour) is outside the range. Please check Draw Time (Hour).
340 112 Host Receiving Text Error Draw Time (Min) is outside the range. Please check Draw Time (Min).
340 113 Host Receiving Text Error Draw Time (Second) is outside the range. Please check Draw Time (Second).
340 114 Host Receiving Text Error Age is outside the range. Please check Age.
340 115 Host Receiving Text Error Age Unit is outside the range. Please check Age Unit.
340 116 Host Receiving Text Error Sex is outside the range. Please check Sex.
340 117 Host Receiving Text Error Cup size is outside the range. Please check Cup size.
340 118 Host Receiving Text Error Sample ID is outside the range. Please check Sample ID.
340 120 Host Receiving Text Error The number of TS exceeds the number of the
regulations.
Please check number of TS.
340 121 Host Receiving Text Error Sample ID is not effective value. Please check Sample ID.
340 122 Host Receiving Text Error Dilution Rate is outside the range. Please check Dilution Rate.
340 123 Host Receiving Text Error Sample Type is not Routine or Stat. Please check Sample Type.
340 124 Host Receiving Text Error Sample ID is outside the range. Please check Sample ID.
340 133 Host Receiving Text Error Application Code is not effective value. Please check Application Code.
340 134 Host Receiving Text Error Dilution Rate is outside the range. Please check Dilution Rate.
341 XX App. Code Conversion
Error. (To HOST)
An application code is not convertible at the
time of HOST communication. * Subcode ”XX”
indicates the application code in [Utility]-
[Application] screen.
Please check whether the application code for HOST is set up correctly.
342 1 App. Code Conversion
Error. (From HOST)
An application code is not convertible at the
time of HOST communication.
Please check whether the application code for HOST is set up correctly.