Section 4
Message Lists
This unit has a self-diagnostic function for internal abnormalities, and an alarm function to display the operation and
alarm.
Abnormality and alarm messages are displayed on the front panel display.
4-1. Description of Message Displays
First line Second line Description Note Solutions
INITIALIZE ERR. 0000 to 0255 Startup error Displays after
5 minutes
from the fail-
ure of startup
Restart the unit. When the error is
occurred many times after restart:
1. Perform the recovery using the
recovery media.
*1
2. Replace the USBDOM, then
perform the recovery using the
recovery media.(“3-9. Proce-
dures After Replacement”)
*1
3. Replace the MB board or Front
board. (Refer to “6-12. MB
Board”, “6-9. Front Board”)
SYSTEM ERR. 0001 Application
communication
error
Displays
when the
Front board
has no
communica-
tion with ap-
plication over
5 minutes.
Restart the unit. When the error is
occurred many times after restart:
1. Perform the “Initialize HDD”.
(Refer to “6-10. HDD”)
2. Perform the “Use Factory Set-
tings”. (Refer to “5-2-4. Use
Factory Settings”)
3. Perform the recovery using the
recovery media.
*1
4. Replace the USBDOM, then
perform the recovery using the
recovery media. (Refer to
“3-9. Procedures After Re-
placement”)
*1
5. Replace the HDD, then per-
form the “Initialize HDD”.
(Refer to “6-10. HDD”,
“5-2-3. Initialize HDD”)
6. Replace the MB board or Front
board. (Refer to “6-12. MB
Board”, “6-9. Front Board”)
Continued
*1: For obtaining the recovery media, please contact your local Sony Sales Office/Service Center.
HVO-500MD/HVO-550MD
4-1