Table 7 (cont)
Characteristics
Performance Requirements
VERTICAL SYSTEM (cont)
Step Response
Rise
Time (0° C
to
+40°
C)
3.5
nanoseconds or less.
Positive-Going Step
(Excluding ADD Mode)
Aberrations (0° C
to
+40°
C)
Position Effect
(0° C
to
+40°
C)
Negative-Going Step
ADD
Mode Operation
76
465B/DM44 Operators
Supplemental Information
5-division reference signal centered
vertically, de coupled
at
all
deflection
factors, from a 25-ohm source with
VAR VOLTS/DIV control
in
calibrated detent position.
+4%,
-4%,4%
p-p or less (5
mV
to
2 V).
+6%,
-6%,
6%
p-p or less (5 V setting
only).
Total aberrations
less
than
+6%,
-6%,
6%
p-p; checked
at
5 mV per division.
Add
2%
to
all
positive-going step
specifications; checked
at
5
mV
per
division.
Add
5%
to
all
aberration specifications;
checked
at
5
mV
per division.
@
I
I
I
I
I
1
I
I
I

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 475A and is the answer not in the manual?

Tektronix 475A Specifications

General IconGeneral
BrandTektronix
Model475A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals