Specifications
This section contains the specifications for the AWG5200 series Arbitrary
Waveform Generators.
All specifications are typical unless noted as warranted. Warranted speci fications
that are m arked with the
symbol are checked in this manual.
Performance conditions
To meet specifications, the following conditions must be met:
The instrument must ha ve been calibrated/adjus ted at an ambient te mperature
between +20 °C and + 30 °C.
The instrument must be operating within the environmental limits. (See
Table 35 on page 24.)
The instrument must be powered from a source that meets the specifications.
(See Table 33 on page 22.)
The instrument must have been operating continuously for at least 20 minutes
within the specified operating temperature range.
AWG5200 Series Technical Reference 1