Troubleshooting
Table 4-7: Calibration error codes (cont.)
Error code Error log message
Possible defect
on board Note
263XY1108
Invalid phase transition distance a t CH2, TEK0015-X
bit Y.
AWG12G AWG12G: U100B, U300B (X=0,1
-> P; X=2,3 -> Q; X=4,5 -> R; X=6,7
-> S; X=8,9 -> T)
270000000
HwExecYigCal PLL read error AWG12G AWG12G: U1000, U1040
270000001
HwExecYigCal PLL 6GHz offset under 2048 vtune=xxxx CLK12G CLK12G:U3200, U3250, Y3280
270000002
HwExecYigCal PLL 6GHz offset underow vtune=xxxx CLK12G CLK12G:U3200, U3250, Y3280
270000003
HwExecYigCal PLL 6GHz tuning underow vtune=xxxx CLK12G CLK12G:U3200, U3250, Y3280
270000004
HwExecYigCal PLL 6GHz tuning underow vtune=xxxx CLK12G CLK12G:U3200, U3250, Y3280
270000005
HwExecYigCal PLL 12GHz tuning over 4095
vtune=xxxx
CLK12G CLK12G:U3200, U3250, Y3280
270000006
HwExecYigCal PLL 12GHz tuning overow vtune=xxxx CLK12G CLK12G:U3200, U3250, Y3280
271000001
HwExecYigCal PLL 6GHz offset under 2048 vtune=xxxx CLK12G CLK12G: U3200, U3250, Y3280
271000002
HwExecYigCal PLL 6GHz offset underow vtune=xxxx CLK12G CLK12G: U3200, U3250, Y3280
271000003
HwExecYigCal PLL 6GHz tuning underow vtune=xxxx CLK12G CLK12G: U3200, U3250, Y3280
271000004
HwExecYigCal PLL 6GHz tuning underow vtune=xxxx CLK12G CLK12G: U3200, U3250, Y3280
271000005
HwExecYigCal PLL 12GHz tuning over 4095
vtune=xxxx
CLK12G CLK12G: U3200, U3250, Y3280
271000006
HwExecYigCal PLL 12GHz tuning overow vtune=xxxx CLK12G CLK12G: U3200, U3250, Y3280
272000000
Clock Power CAL failed
272000001
Clock Power CAL failed at 6 GHz
272000002
Clock Power CAL failed at 12 GHz
290000000
Marker Level Calibration has not been completed. N/A
290000001 Marker cross point calibration result exceeds limit.
OUT10G OUT10G: U120, U170
290X00000
Marker Level Calibration has not been completed at
CHX.
N/A
290XY0001 Marker cross point calibration result exceeds limit at
CHX, MARKER Y.
OUT10G,
AWG12G
OUT10G: U120/170, U470/480,
J100/110/150/170 AWG12G:
J110A/111A/115A/
116A/110B/111B/115B/116B
290XY0002 Marker cross point calibration couldn’t be converged on
-1.0VatCHX,MARKERY.
OUT10G,
AWG12G
OUT10G: U120/170, U470/480,
J100/110/150/170 AWG12G:
J110A/111A/115A/
116A/110B/111B/115B/116B
291000000
Marker Level Calibration has not been completed. N/A
291000001 Marker cross point calibration result exceeds limit.
N/A
291X00000
Marker Level Calibration has not been completed at
CHX.
N/A
AWG7000B and AWG7000C Series Service Manual 4–71

Table of Contents

Other manuals for Tektronix AWG7082C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix AWG7082C and is the answer not in the manual?

Tektronix AWG7082C Specifications

General IconGeneral
BrandTektronix
ModelAWG7082C
CategoryPortable Generator
LanguageEnglish

Related product manuals