Adjustment Procedures
Calibration P
rocedure
Before starting the calibration procedure, you need to restore the calibration
constants that were obtained by ATS (Automatic Test System) at the factory or
the service c
enter.
Restoring the Calibration
Constants
Click the Copy Calibration Data from AWG12G to MIO button in the middle
right of the Service UI window.
The calibration data saved to the EEPROM on the AWG12G board during the ATS
calibration is restored to the EEPROM on the MIO board. The calibration data on
the MIO bo
ard EEPROM is used during the following calibration procedure.
10 MHz Reference
Calibration
This procedure adjusts the built-in TCXO oscillator frequency for the 10 MHz
Referen
ce Output.
Equipment required Prerequisites
One frequency counter
One 50 BNC c able
20 minute warm-up at 20 °C to 30 °C
1. Install the test hookup a nd preset the instrument controls:
a. Connect a BNC cable from the 10 MHz Reference Output on the rear
panel of the arbitrary waveform generator to the CHANNEL 1 of the
freq
uency counter. (See Figure 3-2.)
Figure 3-2: 10 MHz Reference calibration initial test hookup
2. Set the frequency counter CHANNEL 1 Impedance to 50 .
AWG7000B and AWG7000C Series Service Manual 3–5

Table of Contents

Other manuals for Tektronix AWG7082C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix AWG7082C and is the answer not in the manual?

Tektronix AWG7082C Specifications

General IconGeneral
BrandTektronix
ModelAWG7082C
CategoryPortable Generator
LanguageEnglish

Related product manuals