Troubleshooting
Calibration E
rror
To run the instrument calibration, do these steps:
1. Turn off all other applications.
2. From the menu bar, select System and then select Calibration...;the
Calibration dialog box appears.
Figure 4-30: Calibration dialog box
Table 4-7: Calibration error codes
Error code Error log message
Possible defect
on board Note
200XP0001
Analog Level Calibration CHX positive (negative) 5db
attenuator failed. expected 0.56234, actual xxxxx
OUT10G OUT10G: K610, K710
200XP0002
Analog Level Calibration CHX positive (negative) 5db
attenuator failed. expected 0.56234, actual xxxxx
OUT10G OUT10G: K610, K710
200XP0003
Analog Level Calibration CHX positive (negative) 10db
attenuator failed. expected 0.31627, actual xxxxx
OUT10G OUT10G: K620, K720
200XP0004
Analog Level Calibration CHX positive (negative) 20db
attenuator failed. expected 0.100000, actual xxxxx
OUT10G OUT10G: K630, K730
200XP0005
Analog Level Calibration CHX positive (negative)
50mhz filter failed. expected 0.992, actual xxxxx
OUT10G OUT10G: K660, K760
200XP0006
Analog Level Calibration CHX positive (negative)
200mhz filter failed. expected 0.998, actual xxxxx
OUT10G OUT10G: K670, K770
200XP0007
Analog Level Calibration CHX positive (negative)
amplitude cal failed. expected 2.07, actual xxxxx
OUT10G OUT10G: U800
200XP0008
Analog Level Calibration CHX positive (negative) hfd
dac reference failed.
OUT10G AWG10G: U100A/B, U150A/B
200XP0009
Analog Level Calibration CHX positive (negative) output
offset failed. expected 0.50V(0.0V), actual xxxxx
OUT10G OUT10G: U900, U910, U920
200XP0010
Analog Level Calibration CHX internal offset 8bit failed.
expected 0.0mV, actual xxxxxmV
OUT10G AWG10G: U100A/B, U150A/B
4–64 AWG7000B and AWG7000C Series Service Manual