Troubleshooting
Table 4-7: Calibration error codes (cont.)
Error code Error log message
Possible defect
on board Note
200XP0011
Analog Level Calibration CHX i nternal offset 10bit
failed. expected 0 .0mV, actual xxxxxmV
OUT10G AWG10G: U100A/B, U150A/B
200XP0012
Analog Level Calibration CHX can’t nd 8bit outamp
offset cross point. expected 0.0mV, actual pos:
xxxxxmV, neg: xxxxxmV
AWG12G AWG12G: U100A/B, U150A/B
200XP0013
Analog Level Calibration CHX can’t nd 10bit outamp
offset cross point. expected 0.0mV, actual pos:
xxxxxmV, neg: xxxxxmV
AWG12G AWG12G: U100A/B, U150A/B
200XP0015
Analog Level Calibration CHX 5db attenuator gain
balance failed.
200XP0016
Analog Level Calibration CHX 5db attenuator gain
balance failed.
200XP0017
Analog Level Calibration CHX 10db attenuator gain
balance failed.
200XP0018
Analog Level Calibration CHX 20db attenuator gain
balance failed.
200XP0019
Analog Level Calibration CHX lter 50mhz gain balance
failed.
200XP0020
Analog Level Calibration CHX lter 200mhz gain
balance failed.
200XP0021
Analog Level Calibration C HX amp gain balance failed.
210XM0001
Marker Level Calibration CHX MARKER M offset failed. OUT10G OUT10G: U120, U170, U230,
U240, U330, U340
210XM0002
Marker Level Calibration CHX MARKER M positive
offset failed.
OUT10G OUT10G: U120, U170, U230,
U240, U330, U340
210XM0003
Marker Level Calibration C HX MARKER M negative
offset failed.
OUT10G OUT10G: U120, U170, U230, 240,
U330, U340
210XM0004
Marker Level Calibration CHX MARKER M amplitude
failed.
OUT10G OUT10G: U120, U170, U230,
U240, U330, U340
210XM0005
Marker Level Calibration CHX MARKER M amplitude
correction failed.
OUT10G OUT10G: U120, U170, U230,
U240, U330, U340
211XMP006
Marker Level Calibration CHX MARKER M
positive(negative) input level failed.
AWG12G AWG12G: U1000, U100A/B
220000000
HwExecMarkerDelayCal (channel=1) A maximum and a
minimum difference a re smaller than 80ps at coarse=15
AWG12G AWG12G: U1000, U100A
220000001
HwExecMarkerDelayCal (channel=2) A maximum and a
minimum difference a re smaller than 80ps at coarse=15
AWG12G AWG12G: U1000, U100B
220000002
HwExecMarkerDelayCal (channel=1) Frequency
measurement cannot be performed.
AWG12G AWG12G: U1000, U100A
220000003
HwExecMarkerDelayCal (channel=2) Frequency
measurement cannot be performed.
AWG12G AWG12G: U1000, U100B
AWG7000B and AWG7000C Series Service Manual 4–65

Table of Contents

Other manuals for Tektronix AWG7082C