Adjustment Procedures
Required Equipment
You will need the following equipment:
Table 3-2: Test equipment
Description Minimum requirements Examples Quantity
Sampling oscilloscope Bandwidth: 2 0 GHz or higher Tektronix DSA8200 with 80E03
1ea
Oscilloscope Bandwidth: 1 GHz or higher Tektronix TDS5104B
1ea
Frequency counter Accuracy: within ± 0.01 ppm Agilent 53181A 1 ea
50 Ω BNC cable DC to 2 GHz
Tektronix part number
012-0482-00
1ea
50 Ω SMA cable DC to 20 GHz
Tensolite 1-3636-465-5236 2 ea
50 Ω SMA Termination DC to 18 GHz
Tektronix part number
015-1022-01
3ea
SMA-BNC adapter SMA female to BNC male
Tektronix part number
015-0572-00
2ea
50 Ω SMA Attenuator DC to 18 GHz, 12 dB
Hirose AT-112 2 ea
Adjustment Overview
The adjustment p rocedure contains up to ten adjustment items, depending on the
instr
ument type and options. Before performing a djustment p rocedures, you must
warm up the arbitrary waveform generator at least 20 minutes in an ambient
temperature between 20 °C and 30 °C.
Enable the Service Menu
You must enable the Service menu to perform the adjustment procedure. To do
this:
1. Power on the instrument.
2. Se
lect the System menu from the menu bar, and then select Service Mode...
3. The following dialog box is displayed. Enter the password “1185”.
3–2 AWG7000B and AWG7000C Series Service Manual