Commands Listed in Alphabetical Order
REF<x>:HORiz
ontal:DELay:TIMe
Sets or returns the horizontal delay time for reference w aveform <x>, where x
is the reference channe l number. The delay time is expressed in seconds and is
limited to ±5
times the reference horizontal scale.
Group
Vertical
Syntax
REF<x>:HORizontal:DELay:T IMe <NR3>
REF<x>:HORizontal:DELay:T IMe?
Arguments
<NR3> is the delay time in seconds.
Examples
REF2:HORIZONTAL:DELAY:TIM E 4.0E-6 sets the horizontal delay time for
the REF2 waveform to 4 μs.
REF<x>:HORizontal:SCAle
Sets or returns the horizontal scale for reference waveform <x>, where x is the
reference channel number.
Group
Vertical
Syntax
REF<x>:HORizontal:SCAle <NR3>
REF<x>:HORizontal:SCAle?
Arguments
<NR3> is the horizontal scale in seconds.
Examples
REF1:HORIZONTAL:SCALE? might return REF1:HORIZONTAL:SCALE 4.0E-4.
REF<x>:LABel
Sets or returns the reference waveform label for the channel specied by <x>,
where x is the reference channel number.
Group
Vertical
2-230 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual

Other manuals for Tektronix DPO2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO2000 Series and is the answer not in the manual?

Tektronix DPO2000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals