Press the Multiview Zoom front panel button.
Using the multipurpose knobs, set the Zoom position to 25%, and set
the Zoom factor to 20.
Press the Cursors front-panel button once to assign the multipurpose
knobs to the Cursors.
Set the Cursor 1 x position to align the cursor with the nearest trough
of the waveform.
Press the Zoom button. Set the Zoom position to 75%.
Press the Cursors button once. Set the Cursor 2 x position to align
the cursor with the nearest trough of the waveform.
f. Read the measurement:
Press the Clear button. Allow approximately 1000 acquisitions to
accumulate, then press Run/Stop.
Read the Std Dev statistic measurement.
Enter the time on the test record.
The standard deviation (St Dev) measurement must be less than or
equal to the rms Delta-time accuracy limit for your instrument. (See
Table 17: Delta time measurement settings, ≥4 GHz models on
page 250.)
Read the Max and Min statistic measurements.
Subtract the Max and Min statistic measurements.
Enter the time on the test record.
The absolute value of the difference of the Max and Min
measurements must be less than or equal to the pk-pk Delta-time
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
252 MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series

Table of Contents

Other manuals for Tektronix DPO70000C/DX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000C/DX Series and is the answer not in the manual?

Tektronix DPO70000C/DX Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000C/DX Series
CategoryTest Equipment
LanguageEnglish

Related product manuals