9. Calculate the RMS noise (RMS noise = SMV – AMV), and enter the
calculated RMS noise in the test record.
10. Set the Acquisition mode to Sample as follows:
a. Push the front-panel Acquire button.
b. In the Acquisition window, select the Acquisition tab at the left.
c. Select the Sample Acquisition mode.
11. Repeat the tests at 50 Ω as follows:
a. Push the front-panel oscilloscope Vertical V menu button.
b. Set the Termination to 50 Ω.
c. Repeat steps 4 through 10.
12. Repeat the tests at all bandwidth settings as follows:
a. Push the front-panel oscilloscope Vertical V menu button.
b. Set the Termination to 1 MΩ.
c. Select the next Bandwidth at which to test.
d. Starting at step 4, repeat the procedure for each bandwidth setting in the
test record.
13. Press the Menu Off button.
Check sample rate and delay time accuracy
This test checks the sample rate and delay time accuracy (time base).
Performance verification (MSO/DPO5000/B series)
330 MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series

Table of Contents

Other manuals for Tektronix DPO70000C/DX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000C/DX Series and is the answer not in the manual?

Tektronix DPO70000C/DX Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000C/DX Series
CategoryTest Equipment
LanguageEnglish

Related product manuals