Check serial trigger clock recovery range <4 GHz models
Equipment required Equipment required
One precision 50 Ω coaxial cable (Item 4)
One sine-wave generator (Item 9)
One adapter (Item 19)
The oscilloscope must meet the prerequisites.
Also, the instrument must have passed Check
DC Voltage Measurement Accuracy. (See
Signal acquisition system checks on page 176.)
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series 277