Method 2
Equipment required Prerequisites
Sine wave generator (Item 9)
Level meter and power sensor (Item 10)
Two male N to female BNC adapters (Item 14)
Two precision coaxial cables (Item 4)
One or two SMA male-to-female BNC adapters
(Item 19)
(See Prerequisites on page 116.)
1. Install the test hookup: Connect the equipment as shown in the figure below
(start with the sine wave generator connected to the instrument). (See
Figure 47: Equipment setup for maximum amplitude on page 283.)
2. Set the Generator:
Set the sine wave generator to a reference frequency of 10 MHz.
Adjust the sine wave generator amplitude to the required number of
divisions as measured by the instrument.
Figure 47: Equipment setup for maximum amplitude
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series 283