In this section:
Introduction .............................................................................. 2-1
Verification test requirements ................................................... 2-2
Calibration verification limits ..................................................... 2-4
Performing the verification test procedures .............................. 2-5
Front-panel calibration verification ........................................... 2-6
Rear-panel verification ........................................................... 2-46
Introduction
Use the procedures in this section to verify that DMM6500 accuracy is within the limits stated in the
instrument’s one-year accuracy specifications. Specifications and characteristics are subject to
change without notice; refer to the Product Support and Downloads web page
(tek.com/product-support) for the most recent specifications.
You can use these verification procedures to:
• Make sure that the instrument was not damaged during shipment.
• Verify that the instrument meets factory specifications.
• Determine if adjustment is required.
• Verify that adjustment was done properly.
The information in this section is intended for qualified service personnel only, as described
by the types of product users in the Safety precautions pages, provided at the beginning of
this document. Do not attempt these procedures unless you are qualified to do so.
Some of these procedures may expose you to hazardous voltages, that if contacted, could
cause personal injury or death. Use appropriate safety precautions when working with
hazardous voltages.
Section 2
Performance verification