6½-Digit Multimeter with Scanning Calibration and Adjustment Manual Section 3:
DMM6500-905-01 Rev. E June 2021 3-29
ac voltage 22 7 300 kHz
AC current adjustment
The following procedure provides instructions for completing a front-terminal adjustment for ac current.
The following section provides a command parameter table to complete the adjustment.
For this adjustment, you need a:
• Fluke 5720 calibrator
Prepare your DMM6500 for an ac current adjustment
To prepare the DMM6500 for an ac current adjustment:
Connect the Model DMM6500 to the calibrator as shown in the following figure.
Figure 28: Connection for ac current