DMM6500
-Digit Multimeter with Scanning Calibration and Adjustment Manual Section 3:
Adjustment
DMM6500-905-01 Rev. E June 2021 3-29
cal_ACV_700V_1kHz_1pct
ac voltage
22
7
1 kHz
cal_ACV_700V_50kHz_1pct
ac voltage
22
7
50 kHz
cal_ACV_700V_100kHz_1pct
ac voltage
22
7
100 kHz
cal_ACV_700V_200kHz_1pct
ac voltage
22
7
200 kHz
cal_ACV_700V_300kHz_1pct
ac voltage 22 7 300 kHz
cal_ACV_700V_1kHz_fs
ac voltage
1100
750
1 kHz
AC current adjustment
The following procedure provides instructions for completing a front-terminal adjustment for ac current.
The following section provides a command parameter table to complete the adjustment.
For this adjustment, you need a:
Fluke 5720 calibrator
Prepare your DMM6500 for an ac current adjustment
To prepare the DMM6500 for an ac current adjustment:
Connect the Model DMM6500 to the calibrator as shown in the following figure.
Figure 28: Connection for ac current

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY DMM6500 and is the answer not in the manual?

Tektronix KEITHLEY DMM6500 Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY DMM6500
CategoryMultimeter
LanguageEnglish

Related product manuals