Section
2: Performance verification DMM6500 -Digit Multimeter with Scanning
Calibration and Adjustment Manual
2-44 DMM6500-905-01 Rev. E June 2021
Verify the capacitance
Description Verification point Lower limit (F) Upper limit (F)
1 nF range cable REL
0.375
0.25
0.5
10% 1 nF range
0.1
0.0942
0.1058
70% 1 nF range
0.7
0.6894
0.7106
10% 10 nF range
1
0.986
1.014
100% 10 nF range
10
9.95
10.05
10% 100 nF range 10 9.86 10.14
100% 100 nF range
100
99.5
100.5
10% 1 µF range
0.1
0.0986
0.1014
100% 1 µF range
1
0.995
1.005
10% 10 µF range
1
0.986
1.014
100% 10 µF range 10 9.95 10.05
10% 100 µF range
10
9.86
10.14
100% 100 µF range
100
99.5
100.5
Verifying zero values using a 4-wire short
Four-wire short verifications are not included in the Customer Calibration Data Report.
To verify zero values using a 4-wire short, you will:
Check the zero values of various test points with 4-wire connections to the DMM6500 front or rear
terminals.
Verify that the displayed readings are within specified limits.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY DMM6500 and is the answer not in the manual?

Tektronix KEITHLEY DMM6500 Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY DMM6500
CategoryMultimeter
LanguageEnglish

Related product manuals