6½-Digit Multimeter with Scanning Calibration and Adjustment Manual Section 2: Performance v
DMM6500-905-01 Rev. E June 2021 2-37
Verify digitize current 1 A range
Description Calibrator setpoint
(A)
Lower limit Upper limit
Verify digitize current 3 A range
Description Calibrator setpoint
(A)
Lower limit Upper limit
* The 3 A range full-scale test points are limited to 2 A in this table because of the accuracy
limitations of Fluke Models 57xxA and 5725A series calibrators at currents above 2.2 A.
Verify digitize current 10 A range
See Digitize current 10 A range verification (on page 2-49) under Rear-panel verification (on page
2-46).
AC current verification
The following topics describe how to verify ac current.