2: Performance verification DMM6500 -Digit Multimeter with Scanning
Calibration and Adjustment Manual
2-22 DMM6500-905-01 Rev. E June 2021
Equation for 0 °C to 850 °C temperature range
R
RTD
= R
0
(1 + AT + BT
2
)
where:
R
RTD
is the calculated resistance of the RTD
R
0
is the known RTD resistance at 0 °C
T is the temperature in °C
A = alpha [1 + (delta/100)]
B = 1 (alpha)(delta)(1E-4)
The alpha and delta values are listed in the following table.
RTD parameters for equations
The RTD parameters for the Callendar-Van Dusen equations are listed in the following table.
DMM6500 resistance to temperature device (RTD)
Type Standard Alpha Beta Delta R
0
at 0 °C (Ω)
PT100
ITS-90
0.00385055
0.10863
1.49990
100.0000
D100
0.003920
0.10630
1.49710
F100
0.003900
0.11000
1.49589
PT385
IPTS-68
0.003850
0.11100
1.50700
PT3916
0.003916
0.11600
1.50594

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY DMM6500 and is the answer not in the manual?

Tektronix KEITHLEY DMM6500 Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY DMM6500
CategoryMultimeter
LanguageEnglish

Related product manuals