3: Adjustment DMM6500 6½-Digit Multimeter with Scanning
Calibration and Adjustment Manual
3-12 DMM6500-905-01 Rev. E June 2021
Command parameters for a rear-terminal adjustment with a 4-wire
short
When calibrating your DMM6500 for a rear-terminal adjustment with a 4-wire short, use the following
command parameters.
Send each command parameter twice. First, send the parameter using the setup command. Second,
send the parameter using the execute command.
The example code below uses the first line of code from the table below. Send all command
parameters using this command.
cal.adjust.step.setup("cal_DCV_100mV_zero_rear_setup")
cal.adjust.step.execute("cal_DCV_100mV_zero_rear_setup")
Command parameters
cal_3W_1kohm_SLO_zero_rear
cal_3W_10kohm_SLO_zero_rear
cal_3W_100kohm_SLO_zero_rear