Performance Ver
ification
21. Turn the Multipurpose b knob to select the input channel to be tested.
22. Push Menu Off on the front panel. The measured value should appear in a m easurement pane at the top of the display.
23. E nter the measured value in the test record. (See page 67, DVM Voltage Accuracy Tests (DC).)
24. Repeat the procedure (steps 6, 7, 8 and 22) for each volts/division s etting shown in the test record.
25. Repeat all steps, starting with step 1, for each oscilloscope channel you want to check.
This completes the procedure.
MDO3000 Series S pecifications and Performance Verification 115