Performance Ver
ification
Level Measurement Uncertainty Tests
Table 30: Level Measurement Uncertainty
Level Measurement Uncertainty
Performance checks
Low limit Test result High limit
9 kHz –1.2 dB +1.2 dB
50 kHz –1.2 dB +1.2 dB
100 kHz – 900 kHz –1.2 dB +1.2 dB
1MHz–9MHz –1.2dB +1.2dB
10 MHz - 90 MHz –1.2 dB +1.2 dB
+10 dBm All models
100 MHz – B W –1.2 dB +1.2 dB
9 kHz –1.2 dB +1.2 dB
50 kHz –1.2 dB +1.2 dB
100 kHz – 900 kHz –1.2 dB +1.2 dB
1MHz–9MH
z
–1.2 dB +1.2 dB
10 MHz - 90 MHz –1.2 dB +1.2 dB
0 dBm All models
100 MHz – B W –1.2 dB +1.2 dB
9 kHz –1.2 dB +1.2 dB
50 kHz –1.2 dB +1.2 dB
100 kHz – 900 kHz –1.2 dB +1.2 dB
1MHz–9
MHz
–1.2 dB +1.2 dB
10 MHz – 90 MHz –1.2 dB +1.2 dB
–15 dBm All m od
els
100 MHz – B W –1.2 dB +1.2 dB
Func
tional che ck with a TPA -N-PRE Preamp Attached
Table 31: Functional check with a TPA-N-PRE Preamp attached
Functional check with a TPA-N-PRE Preamp attached
Performance checks
Limit Test result
1.7 GHz
1.5 dB
All models
2.9 GHz
1.5 dB
64 MDO3000 Series Specifications and Performance Verification