Commands Listed in Alphabetical Order
Examples
WFMOUTPRE:XZE
RO?
might return :WFMOUTPRE:XZERO 5.6300E-9 indicating
that the trigger occurred 5.63 ns before the first sample in the waveform record.
WFMOutpre:YMUlt? (Query Only)
This query returns the vertical scale multiplying factor used to convert
the waveform data point values in the outgoing waveform record from
digitizing levels to the YUNit units. You can determine the units by using the
WFMOutpre
:YUNit
query.
See the description of the WFMInpre:YMUlt command to see how this scale
factor is
used to convert waveform sample values to volts.
If the waveform specified by DATa:SOUrce is not turned on, nothing is returned
andaner
ror event is generated.
(See page 2-95, Waveform Transfer Command Group.)
(See page D-1, Waveform Transfer (WFMOutpre and CURVe Query) Examples.)
Group
Waveform Transfer
Syntax
WFMOutpre:YMUlt?
Related Commands
DATa:SOUrce,
WFMInpre:YMUlt
Examples
WFMOUTPRE:YMULT? might return :WFMOUTPR E:YMULT 4.0000E-3 indicating
that the vertical scale for the corresponding waveform is 100 mV/div (for 8-bit
waveform data or 0.390625 V/div for 16–bit data).
WFMOutpre:YOFf? (Query O nly)
This query returns the vertical position of the source waveform in digitizing
levels. There are 25 digitizing levels per vertical division for 1-byte data, and
6400 digitizing levels per vertical division for 2-byte data. See the description of
WFMInpre:YOFf to see how this position is used to convert waveform sample
values to volts. If the waveform specified by DATa:SOUrce is not turned on,
nothing is returned and an error event is generated.
(See page 2-95, Waveform Transfer Command Group.)
(See page D-1, Waveform Transfer (WFMOutpre and CURVe Query) Examples.)
2-972 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual