Commands Listed in Alphabetical Order
MARKER:M<x>:
NOISEDensity? (Query Only)
This command returns the noise density of the RF_NORMal trace at the specified
marker position in <RF Units>/Hz units, where <RF Units> are the units specified
by the comman
d RF:UNIts.
Conditions
MDO3000, MDO4000/B, or MDO4000C models with option SA3 or SA6 only.
Group
RF
Syntax
MARKER:M<x>:NOISEDensity?
Related Commands
MARKER:MANual, MARKER:M<x>:PHASENoise?
Returns
floating point number
Examples
MARKER:M2:NOISEDensity? might return –133.2350, indicating noise density
of –133.2350 dBm/Hz at the marker position.
MARKER:M<x>:PHASENoise? (Query Only)
This command returns the phase noise of the RF_NORMal trace at the specified
marker position in dBc/Hz units.
Conditions
MDO3000, MDO4000/B, or MDO4000C models with option SA3 or SA6 only.
Group
RF
Syntax
MARKER:M<x>:PHASENoise?
Related Commands
MARKER:MANual, MARKER:M<x>:NOISEDensity?
Returns
floating point number
Examples
MARKER:M2:PHASENoise? might return –60.1289, indicating phase noise of
–60.1289 dBc/Hz at the marker position.
2-362 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual