Commands Listed in Alphabetical Order
Group
Measurement
Syntax
MEASUrement:IMMed:DELay:DIRection {BACKWards|FO RWards}
MEASUrement:IMMed:DELay:DIRection?
Related Commands
MEASUrement:IMMed:SOUrce<x>
Arguments
BACKWards starts the search at the end of the waveform and looks for the last
rising or falling edge in the waveform.
FORWards starts the search at the beginning of the waveform and looks for the
first rising or falling edge in the waveform.
Examples
MEASUREMENT:IMMED:DELAY:DIRECTION FORWARD S starts searching from the
beginning of the waveform record and looks for the first rising or falling edge.
MEASUREMENT:IMMED:DELAY:DIRECTION? might return
:MEASUREMENT:IMMED:DELAY:DIRECTION BACKWARDS indicating
that searching begins at the end of the waveform record and looks for the last
rising or falling edge.
MEASUrement:IMMed:DELay:EDGE<x>
This command specifies the slope of the edge the oscilloscope uses for the delay
"from" or "to" waveform when taking an immediate delay measurement.
Group
Measurement
Syntax
MEASUrement:IMMed:DELay:EDGE<x> {FALL|RISe}
MEASUrement:IMMed:DELay:EDGE<x>?
Related Commands
MEASUrement:IMMed:SOUrce<x>
Arguments
<x> specifies which waveform to use, where <x> = 1 is the "from" waveform, and
<x> = 2 is the "to" waveform.
FALL specifies the falling edge.
RISe specifies the rising edge.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-423