EasyManuals Logo
Home>Tektronix>Test Equipment>MDO3000 Series

Tektronix MDO3000 Series Program Manual

Tektronix MDO3000 Series
1158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #496 background imageLoading...
Page #496 background image
Commands Listed in Alphabetical Order
Group
Power
Syntax
POWer:GATESOurce {CH1|CH2|CH3|CH4|REF1|REF2|REF 3|REF4|NONe}
POWer:GATESOurce?
Arguments
CH1–CH4 sets an analog channel as the gate source.
REF1–REF4 sets a reference waveform as the gate source.
NONe is set when the gate source is not used in the application.
Examples
POWER:GATESOURCE CH1 sets the gated source as CH1.
POWER:GATESOURCE? might return POWER:GATESOURCE CH2 indicating that
CH2 is the gated source.
POWer:GATing
This command specifies the power application gating.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:GATing {OFF|SCREen|CURSor}
POWer:GATing?
Arguments
OFF turns off measurement gating (full record).
SCREen turns on gating, using the left and right edges of the screen.
CURSor limits measurements to the portion of the waveform between the vertical
bar cursors, even if they are off screen.
Examples
POWER:GATING CURSOr sets cursor gating on the oscilloscope.
POWER:GATING? might return POWER:GAT ING OFF indicating that gating is
turned off on the oscilloscope.
2-466 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Table of Contents

Other manuals for Tektronix MDO3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO3000 Series and is the answer not in the manual?

Tektronix MDO3000 Series Specifications

General IconGeneral
BrandTektronix
ModelMDO3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals