Commands Listed in Alphabetical Order
Conditions
This command re
quires a DPO4AUTO or DPO4AUTOMAX application module
on DPO/MSO4000B and MDO4000/B/C models and a MDO3AUTO application
module on MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CA N:DATa:DIRection
{READ|WRITE|NOCARE}
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CA N:DATa:DIRection?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:CONDition
Arguments
READ specifies the read direction.
WRITE specifies the write direction.
NOCARE specifies either a read or write direction.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:DATa:OFFSet
This command specifies the data offset in bytes to search for in a CAN search.
The default value is 0. SEARCH<x> is the search
number, which is always 1, and
B<x> is the bus number (1-4). 1-3 for the MDO4000C and 1-2 for the MDO3000.
Conditions
This command requires a DPO4AUTO or DPO4AUTOMAX application
module on a DPO/MSO4000B and MDO4000/B/C models and a MDO3AUTO
application module on MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CA N:DATa:OFFSet <NR1>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CA N:DATa:OFFSet?
Related Commands
BUS:B<x>:CAN:STANDard
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:CONDition
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:DATa:SIZe
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:DATa:VALue
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-625