Trigger Setup
Selecting Triggers
Trigger Type Trigger Conditions
Edge
Trigger on a rising edge, a falling edge, or both edges,
as dened by t
he slope control. Coupling c hoices are
DC, LF Reject, HF Reject, and Noise R eject.
Edge triggers are the simplest and most commonly
used trigge
r type, with both analog and digital signals.
An edge trigger event occurs when the trigger source
passes through a specied voltage level in the specied
direction.
Sequence
(B Trigger) (not
available with an
edge slope
of
Both)
Combine an
edge A Event (Main) trigger with the B
Event (Delayed) trigger to capture more complex signals.
(See page 65, Act on Event.)
Time. Aft
er the A Event occurs, the trigger system waits
the specied amount of time, and then looks for the B
Event before triggering and displaying the waveform.
Events. After the A Event occurs, the trigger system
looks fo
r a specied number of B Events before
triggering and displaying the w aveform.
Pulse Width Trigger on pulses that are less than, greater than, equal
to, or no
t equal to a specied time. Additionally, you can
trigger when a pulse width is within or outside a range of
two different specied times. You can trigger on positive
or nega
tive pulses. Pulse width triggers a re primarily
used on digital signals.
Timeo
ut
Trigger when no pulse is detected within a specied
time. The signal stays above or below (or either above
or bel
ow) a set value for a set amount of time.
Runt Trigger on a pulse amplitude that crosses one threshold
but f
ails to cross a second threshold before recrossing
the rst. You can detect positive or negative (or either)
runts, or only those wider than, less than, greater than,
equ
al to, or not equal to a specied width. Runt triggers
are primarily used on digital signals.
MDO3000 Series Oscilloscopes User Manual 93

Table of Contents

Other manuals for Tektronix MDO3104

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO3104 and is the answer not in the manual?

Tektronix MDO3104 Specifications

General IconGeneral
BrandTektronix
ModelMDO3104
CategoryTest Equipment
LanguageEnglish

Related product manuals