Trigger Setup
Starting and Stopping an Acquisition
After y ou have dened the acquisition and trigger parameters, start the acquisition with Run/Stop or Single.
Push Run/Stop to start a cquisitions.
The oscilloscope acquires repeatedly
until you push the button again to stop
the acquisition.
Push Single to take a single acquisition.
Single sets the trigger mode to Norm al
for the single acquisition.
If any analog or digital waveforms are
active, then pressing Run/Stop to stop
acquisitions will cause the oscilloscope
to wait for one more trigger event before
stopping. O nce the acquisition occurs,
the Run/Stop button will turn red and the
Single button will become unlit.
If the Trigger Mode is set to Auto and
another trigger event does not occur
within the Auto trigger timeout period
then an acquisition will be made and the
instrument will stop.
If the Trigger mode is set to Normal,
the oscilloscope will continue waiting
for a trigger event to occur as long as
necessary.
102 MDO3000 Series Oscilloscopes User Manual

Table of Contents

Other manuals for Tektronix MDO3104

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO3104 and is the answer not in the manual?

Tektronix MDO3104 Specifications

General IconGeneral
BrandTektronix
ModelMDO3104
CategoryTest Equipment
LanguageEnglish

Related product manuals