Commands Listed in Alphabetical Order
Examples
AFG:RAMP:SYMM
ETRY 50.0
sets the AFG ramp symmetry to 50 percent.
AFG:SQUare:DUty
Sets (or queries) the AFG duty cycle, as a percentage. The minimum is 10.0%,
maximum is 90
.0% and increment is 0.10%.
Conditions
MDO3000 series with option MDO3AFG installed and MDO4000C series with
option MDO4AFG installed only.
Group
AFG
Syntax
AFG:SQUare:DUty <NR3>
AFG:SQUare:DUty?
Arguments
Floating point number that represents the AFG duty cycle, as a percentage.
Examples
AFG:SQUARE:DUTY 50.0 sets the AFG duty cycle to 50 percent.
ALIas:CATalog? (Query Only)
Returns a list of the currently defined alias labels, separated by commas. Ifno
aliases are defined, the query returns the string "".
Group
Alias
Syntax
ALIas:CATalog?
Examples
ALIAS:CATALOG? might return the string :AL IAS:CATALOG
"SETUP1","TESTMENU1","DEFAULT"
showing that there are three aliases
named SETUP1, TESTMENU1, and DEFAULT.
ALIas:DEFine
Assigns a sequence of program messages to an alias label. These messages are
then substituted for the alias whenever it is received as a command or query,
provided that ALIas:STATE has been turned on. The query form of this command
returns the definitions of a selected alias.
2-142 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual