Commands Listed in Alphabetical Order
HORizontal:S
CAle
This command specifies the time base horizontal scale.
Group
Horizontal
Syntax
HORizontal:SCAle <NR3>
HORizontal:SCAle?
Arguments
<NR3> is a floating point number that specifies the range from 400 ps to 1000 s,
depending on the record length.
Examples
HORIZONTAL:SCALE 2E-6 sets the main scale to 2µs per division.
HORIZONTAL:SCALE? might return :HORIZONTAL:SCALE 2.0000E-06
indicating that the main scale is currently set to 2 µs per division.
ID? (Query Only)
Returns the instrument identification data. This is similar to the data returned by
the *IDN? IEEE488.2 common query, but also includes any enabled application
modules and enabled options. However, it does not include the instrument
serial number.
The format of the returned string is an extension of that specified in Tektronix
Codes and Formats with the addition of the application information (and is not
dependent upon :HEADer or :VERBose settings):
ID TEK/<model>,CF:91.1CT FV:v<firmware version> <app module info>
where:
<firmware version> is comprised of 2 numerical values separated by a decimal
point representing the major and minor version numbers, respectively.
<app module info> is a list of application modules separated by a space of
the form:
<App ID> such as DPO4EMBD:V1.00, MDO3AUTO:V1.00, etc.
Group
Miscellaneous
Syntax
ID?