Commands Listed in Alphabetical Order
RF:DETECTionm
ethod:RF_MINHold?
Related Commands
RF:DETECTionmethod:MODe, RF:DETECTionmethod:RF_AVErage,
RF:DETECTionmethod:RF_MAXHold, RF:DETECTionmethod:RF_NORMal
Arguments
PLUSpeak — Detection method that selects the largest value in each interval.
MINUSpeak — Detection method that selects the smallest value in each interval.
SAMple — Detection method that selects the first point in each interval.
AVErage — Detection method that selects the mean value within an interval.
Examples
RF:DETECTionmethod:RF_MINHold SAM changes the detection method to
sample.
RF:DETECTionmethod:RF_MINHold? might return AVERAGE, indicating the
detection method has been set to select the mean value within a waveform interval.
RF:DETECTionmethod:RF_NORMal
This command specifies the detection method the oscilloscope should use when
acquiring an RF Normal frequency domain trace. The Normal trace displays the
most recently acquired sample at each trace point. .
In order to change the detection method, you must first change the detection
method mode to
MANUAL using the command RF:DETECTionmethod:MODe.
The detection method is the internal calculation that decimates (reduces)a
long frequency domain waveform down to display resolution, by dividing the
waveform into intervals and choosing a single value to represent each interval
on screen.
Conditions
MDO3000, MDO4000/B, or MDO4000C models with option SA3 or SA6 only.
Group
RF
Syntax
RF:DETECTionmethod:RF_NORMal
{PLUSpeak|MINUSpeak|SAMple|AVErage}
RF:DETECTionmethod:RF_NORMal?
Related Commands
RF:DETECTionmethod:MODe, RF:DETECTionmethod:RF_AVErage,
RF:DETECTionmethod:RF_MAXHold, RF:DETECTionmethod:RF_MINHold
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-549