Ta ble 4: Trigger specications (cont.)
MIL-STD-1553 Bit Rate: 1Mb/s
Trigger on: Sync, Word Type (Command, Status, Data), Command Word (set RT Address
(=, ,<,>,, , inside range, outside range), T/R, Sub-address/Mode, Data Word
Count/Mode Code, and Parity individually), Status Word (set RT Address ((=, ,<,>,,
, inside range, outside range), Message Error, Instrumentation, Service Request Bit,
Broadcast Command Received, Busy, Subsystem Flag, Dynamic Bus Control Acceptance
(DBCA), Terminal Flag, and Parity individually) Data Word (user-specied 16-bit data value)
Error (Sync, Parity, Manchester, Non-contiguous data) Idle Time (minimum time selectable
from 4 s to 100 s; maximum time selectable from 12 s to 100 s; trigger on < minimum,
> m aximum, inside range, outside range)
USB Data Rates Supported: HS: 480 Mbps, Full: 12 Mbps, Low: 1.5 Mbps
Trigger On: Sync, Reset, Suspend, Resume, End of Packet, Token (Address) Packet, Data
Packet, Handshake Packet, Special Packet, Error
NOTE. HIGH SPEED support available only on MDO4104-3 and MDO4104-6 models.
Ethernet
Bit Rate: 10BASE-T, 10 Mbps; 100BASE-TX, 100 Mbps
Trigger On: Start Frame Delimiter (SFD), MAC Address, MAC Length/Type, IP Header, TCP
Header, TCP/IPv4/MAC Client Data, End of Packet, Idle, FCS (CRC) Error, MA C Q-Tag
control Information.
Digital Acquisition System Specications
The following table shows the digital acquisition specications for the MDO4000
Series oscilloscopes.
Ta ble 5: Digital acquisition specications
Characteristic Description
Threshold voltage range –40 V to +40 V
Digital channel timing resolution
2 ns main memory, 60.6 ps for MagniVu memory
Logic threshold accuracy
±(100 mV + 3% of threshold setting after calibration)
Requires v alid SPC, as described in step 2 of the Self Test. (See page 62, Self Test.)
Minimum detectable pulse width,
typical
1ns
Using M agniVu memory. Requires the use of 342-1140-00 ground clip on each channel.
MDO4000 Series Specications and Performance Verication 15

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO4104 and is the answer not in the manual?

Tektronix MDO4104 Specifications

General IconGeneral
BrandTektronix
ModelMDO4104
CategoryTest Equipment
LanguageEnglish

Related product manuals