Specications
Frequency
RF input frequency range 9kHzto6.2GHz
Frequency reference accuracy
Over 18 °C to 28 °C ambient
temperature range
±3 x 10
–6
+ aging (after 20 minute warm up)
Aging, typical
±3 x 10
–6
(rst year), ±1 x 10
–6
each year thereafter
Over operating temperature
range (–10
°C to +55 °C
ambient), typical
±25 x 10
–6
+ aging
External frequency reference
input
Input Frequency Range 10 MHz ±10 Hz
Input Lev
el Range
–10 dBm to
+10 dBm sinusoid
Impedance
50
Center
frequency resolution
Block I
Q samples
1Hz
Strea
med ADC samples
500 kHz