Performance ver
ication
Test record
Print out the following test record pages and use them to record the performance test results for the RSA306 spectrum
analyzer.
Table 26: Test record: RSA306
Instrument Serial Number:
Temperature:
Date of Calibration:
Certicate Number:
RH %:
Technician:
Frequency accuracy
Measured Frequency
(1 GHz CF)
Instability measured,
ppm [(Marker
Frequency
1E9)/1000]
Total specied
Instability (aging
plus other drift) Pass/Fail
External reference
Phaselock to 10 MHz,
0 dBm Pass/Fail
Amplitude accuracy
Reference level Error, dB Specication, dB P ass/Fail
+20 dBm < 3 GHz
±2.0 dB
+20 dBm > 3 GHz
±2.75 dB
+0 dBm < 3 GHz
±2.0 dB
+0 dBm > 3 GHz
±2.75 dB
–13 dBm < 3 GHz
±2.0 dB
–13 dBm > 3 GHz
±2.75 dB
–30 dBm < 3 GHz
±2.0 dB
–30 dBm > 3 GHz
±2.75 dB
Channel amplitud e atness
Center Frequency Peak Error, dB Specication, dB Pass/Fail
22 MHz < ±1 dB
1280 MHz < ±1 dB
2160 MHz < ±1 dB
RSA306 Specications and Performance Verication 54

Other manuals for Tektronix RSA306

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix RSA306 and is the answer not in the manual?

Tektronix RSA306 Specifications

General IconGeneral
BrandTektronix
ModelRSA306
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals