Performance ver
ication
Table 2: RSA306 required test equipment (cont.)
Item Description Qty Model Number Purpose
Band pass lter 2150 MHz, SMA
2
Mini-Circuits Labs
ZX75BP-2150+
For third-order distortion
measurement
Low pass lter 3300 MHz cutoff, L250 lter
1
K&L 5L3-3300/E 10000
O/OP
Spurious test
Tunable lter
375 MHz to 750 MHz, N
connectors
1
K&L 5BT-375/750-5-N/N Spurious test
Tunable lter
750 MHz to 1500 MHz, N
connectors
1
K&L 5BT-750/1500-5-N/N Spurious test
Tunable lter
1500 MHz to 3000 MHz, N
connectors
1
K&L 5B T-1500/3000-5-N/N Spurious test
Cable, SMA-to-N ST18/SMAm/Nm/36in
1
Huber-Suhner 84004594 Generator output to RF
input
Cable, SMA ST18/SMAm/SMAm/36in
2
Huber-Suhner 84002061 TOI and tests requiring
low-pass lters
Cable, SMA ST18/SMAm/SMAm/8
2
Huber-Suhner 84028563 TOI tests
Cable, BNC-to-SMA BNC(m) to SMA(m) cable -
1 meter
1 Pasternack PE3615-36
Generator timebase output
to Ref IN.
(External-timebase to
generator-timebase input
may need a similar cable)
Torque wrench 12 in-lb - Type N 1
Maury Microwave 2698C2
N- connector attachments
Torque wrench 8 in-lb Torque wrench - 3.5 mm 1
Huber-Suhner 74Z-0-0-21 SMA-connector
attachments
NOTE. Make sure that any adaptor and cable you use is specied to operate a t the frequency range of the test you
are performing. Connector frequency ranges:
BNC:DCto1GHztyp
ical, up to 3 GHz for certain BNC cable/connectors.
N:DCto18GHztypical
SMA: DC to 18 GHz typical, up to 26.5 GHz for certa
in SMA cable/connectors.
Preliminary c hecks
Do these steps before starting the performance verication procedures.
Warm up the instrument
1. Connect the RSA306 USB cable to the host PC.
The LED on the RSA306 should initially glow red then turn green
after a few moments.
2. Make sure the SignalVu-PC application is connected to the RSA 306 over USB.
RSA306 Specications and Performance Verication 12

Other manuals for Tektronix RSA306

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix RSA306 and is the answer not in the manual?

Tektronix RSA306 Specifications

General IconGeneral
BrandTektronix
ModelRSA306
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals