Performance ver
ication
External reference input, functional test
This check is a functional check. It is an important check for customer use, but does not check warranted speci cation lim its .
1. Connect a signal generator RF output to the N-connector Ref input of the R SA306, as shown. The signal generator
accuracy must be better than ±0.05 ppm. If it does not, it must have its frequency reference phase locked to a precision
frequency reference.
Note: The Stanford Research Systems SG386 signal generator has adequate frequency stability w ithout use o f an
external timebase, but only if it has been calibrated/veried within 1 year. If this is not the case, the signal generator must
have an accurate external timebase connected to its timebase input (10 M Hz).
2. Set the signal generator controls:
a. Frequency = 10 MHz
b. Amplitude =0dBm
3. Set the RSA306 to use the external reference (Setup > Acquire > Frequency Reference).
4. Under the source eld, select the External (10 MHz) radio button.
5. Check that the Status Bar shows Ref: Ext. This generally occurs within 5 seconds.
6. Enter pass/fail result in the test record.
15 RSA306 Speci cations and Performance Verication

Other manuals for Tektronix RSA306

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix RSA306 and is the answer not in the manual?

Tektronix RSA306 Specifications

General IconGeneral
BrandTektronix
ModelRSA306
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals