Trigger Contr
ols
Trigger Menu. When it is pressed once, it d isplays the Trigger Menu. When it is
kept pressed for more than 1.5 seconds, it will show the trigger view, meaning
it will di
splay the trigger waveform in place of the channel waveform. Use the
trigger view to see how the trigger settings, such as c oupling, affect the trigger
signal. Releasing the button will stop the trigger view.
Level. When you use an Edge or Pulse trigger, the Level knob sets the amplitude
level that the signal must cross to acquire a waveform. Push this knob to set
the tri
gger level to the vertical midpoint between the peaks of the trigger signal
(set to 50%).
Force Trig. Use this to complete the waveform acquisition whether or not the
oscilloscope detects a trigger. This is useful for single sequence acquisitions
and Normal trigger mode. (In Auto trigger mode, the oscilloscope automatically
forc
es triggers periodically if it does not detect a trigger.)
Menu and Control Buttons
TBS1000B and TBS1000B-EDU Series Oscilloscopes User Manual 13