TDS 500B, 600B and TDS 700A Service Manual
4–15
Performance Tests
This section contains a collection of procedures for checking that the TDS 500B,
600B, and 700A Digitizing Oscilloscopes perform as warranted.
The procedures are arranged in four logical groupings: Signal Acquisition System
Checks, Time Base System Checks, Triggering System Checks, and Output Ports
Checks. They check all the characteristics that are designated as checked in
Specifications. (The characteristics that are checked appear in boldface type
under Warranted Characteristics in Specifications.)
STOP. These procedures extend the confidence level provided by the basic
procedures described on page 4–5. The basic procedures should be done first,
then these procedures performed if desired.
Prerequisites
The tests in this section comprise an extensive, valid confirmation of perform-
ance and functionality when the following requirements are met:
H The cabinet must be installed on the Digitizing Oscilloscope.
H You must have performed and passed the procedures under Self Tests, found
on page 4–5, and those under Functional Tests, found on page 4–7.
H A signal-path compensation must have been done within the recommended
calibration interval and at a temperature within ±5_ C of the present
operating temperature. (If at the time you did the prerequisite Self Tests, the
temperature was within the limits just stated, consider this prerequisite met.)
H The Digitizing Oscilloscope must have been last adjusted at an ambient
temperature between +20_ C and +30_ C, must have been operating for a
warm-up period of at least 20 minutes, and must be operating at an ambient
temperature between +4_ C and either +45_ C for the TDS 600B or +50_ C
for the TDS 500B and 700A. (The warm-up requirement is usually met in
the course of meeting the Self Tests and Functional Tests prerequisites listed
above.)

Table of Contents

Other manuals for Tektronix TDS 700A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 700A and is the answer not in the manual?

Tektronix TDS 700A Specifications

General IconGeneral
BrandTektronix
ModelTDS 700A
CategoryTest Equipment
LanguageEnglish

Related product manuals