Maintenance
6-- 30
TDS1000 and TDS2000 Series Digital Storage Oscilloscopes Service Manual
You will need a torque-limiting Torx T-15 screwdriver for this procedure.
Removal. To remove the front case, follow these steps:
1. Remove the power button and rear case using the procedures on page 6--10.
2. Remove the entire internal assembly from the front case using the procedure
on page 6--17.
3. Remove the switch keypad using the procedure on page 6--28.
Installation. To install the front case, refer to Figure 6--9 on page 6--17 and follow
these steps:
1. Place the switch keypad into the front case.
2. Place the entire internal assembly into the front case.
3. Use the installation procedure for each module removed to reassemble the
oscilloscope.
Front Case