TDS1000 and TDS2000 S eries Digital Storage Oscilloscopes Service Manual
4-1
Performance Verification
This chapter contains performance verification procedures for the specifications
marked with the n symbol. The following equipment, or a suitable equivalent,
is required to complete these procedures.
Required Equipment
Description
Minimum
requirements
Examples
DC Voltage Source 17.5mVto7V,±0.5%
accuracy
Wavetek 9100 Universal
Calibration System with
Os
illos
o
eC
libr
tion
Leveled Sine Wave Generator 50 kHz and 200 MHz, ±3%
amplitude accuracy
Module (Option 250)
Fluke 5500A Multi-product
Time Mark Generator 10 ms period, ±10 ppm
accuracy
C
libr
tor with Os
illos
ope
Calibration Option (Option
5500A-SC)
50 Ω BNC Cable BNCmaletoBNCmale,
≈ 1 m (36 in) long
Tektronix part number
012-0482-XX
50 Ω BNC Cable BNCmaletoBNCmale,
≈ 25 cm (10 in) long
Tektronix part number
012-0208-XX
50 Ω Feedthrough
Termination
BNC male and female
connectors
Tektronix part number
011-0049-XX
Dual Banana to BNC Adapter Banana plugs to BNC female Tektronix part number
103-0090-XX
BNC T Adapter BNC male to dual BNC
female connectors
Tektronix part number
103-0030-XX
Splitter, Power Frequency range: DC to
4 GHz. Tracking: >2.0%
Tektronix part number
015--0565--XX
Adapter (four required) Male N--to --female BNC Tektronix part number
103--0045--XX
Adapter Female N --to --male BNC Tektronix part number
103--0058--XX