When the Signal Path
compensation test status is Pass,
continue calibrating the probe.
See Compensate the probe
on page 14.
Compensate the probe
Connect a BNC cable from the DC
Probe Cal output connector on
the oscilloscope to the BNC
connector on the TriMode DC
Calibration board.
Connect the cable from the
TekConnect TekVPI adapter to the
connector on the calibration
board.
Connect the other end of the
TekConnect TekVPI adapter cable
to the Aux In channel on the
oscilloscope.
Connect the TEK-CDA accessory
to an oscilloscope channel. ????
Connect the probe to any channel
of the oscilloscope. Allow the
probe to warm up for 20 minutes.
Set the oscilloscope to display
the channel.
Depending on your setups, do
one of the following:
Functional check and calibration
P7700 Series TriMode™ Probes 14