display is achieved only when the correct trigger mode is selected.
Try changing the trigger coupling to High Frequency Holdoff or Low
Frequency Holdoff or to filter any high or low frequency noise that is interfering with
triggering.
(4) Slow refresh
. Check if the acquiring mode in ACQUIRE button menu is average, and
the average times are large.
. If increasing the refresh speed is needed, please decrease the average
times appropriately or select other acquiring modes, i.e. normal sampling.
. Check if the persistence in DISPLAY button menu is set relatively long or
infinite.
(5) The waveform is like stair-step shape.
. Stair-step waveform is normal. The horizontal time-base scale may be is
overlow, increase the horizontal time base to increase the horizontal
resolutuion, which can improve the display.
. The display type may be vector, the connecting line between sampling
points may lead to a stair-step shape. Set the display type as point display
to solve the problem.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the UNI-T UTD2000 Series and is the answer not in the manual?

UNI-T UTD2000 Series Specifications

General IconGeneral
BrandUNI-T
ModelUTD2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals