c) Oxygen saturation in the air
Follow the steps below to take a measurement:
Connect the probe to the meter and turn on the meter.
Remove the red protective cap (21) from the membrane head.
To switch the measurement function to “O2”, press the “▼ FUNCTION” button (8). The
switchoverisconrmedwithabeep.The“%O2”unitwillappearonthedisplay.
Leave the probe in the air for some time. During this time, the sensor can adapt to the tem-
perature.
During the measurement, the probe must be moved around by the air (at least 0.2 m/s). This
can be achieved by permanent movement of the probe.
The upper display line shows the measured value of oxygen saturation (%O2).
The lower display line shows the measured temperature value of the measured ambient air.
Afteryounishtesting,alwaysswitchthemeteroff.Replacetheprobeprotectivecap.
d) Holding the measured value (“HOLD”)
The current measured value can be held to obtain a longer reading time. To hold the measured
value, press the “HOLD” button.
The keystroke is signalled with a beep. The measured value and the “HOLD” symbol appear
on the display.
To disable the function, press the “HOLD” button again. “HOLD” disappears from the display.
39

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the VOLTCRAFT DO-101 and is the answer not in the manual?

VOLTCRAFT DO-101 Specifications

General IconGeneral
BrandVOLTCRAFT
ModelDO-101
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals