Service and maintenance toolset
Connex® Spot Monitor
The following is the listing of tools and test equipment required to perform basic
functional verification checks, full functional verification checks and calibration, and
service and maintenance on the Welch Allyn Connex Spot Monitor. Please refer to the
device service manual for detailed instructions to service your device.
Note Tools and test equipment must also be readily available to complete Welch
Allyn Technical Service Training for this device.
Material no. Description Qty
Usage
Full Basic
General test equipment
106270 Service test box 1 X
106275 Micro USB cable 1 X
N/A Wireless router, a/b/g or equivalent 1 X
N/A Ethernet cable 1 X
N/A PC running Windows XP SP3, Windows 7, or Windows 8 1 X X
web download Welch Allyn Service Tool 1 X X
1001120 Welch Allyn Service Tool, Gold license 1 X
NIBP testing
407672 BP test volume repair fixture 113670 1 X X
N/A Pressure meter (must include at least two decimal points and be
accurate to within ±0.5 mmHg)
1 X X
N/A Blood pressure Y-tube 1 X X
6000-30 Single tube blood pressure hose, 5 feet 1 X X