Disassemble the front housing
Remove the bezel from the front housing
Lift the rubber bezel up and over the power switch cable and the 6 bosses on the
housing.
Reassembly notes
Kit items:
105922, Bezel
105921, Front housing (housing, power switch, and 6 Phillips screws)
Note Reassembly is not the reverse of assembly.
Attach the bezel to the front housing
1. Align the bezel with the front housing and feed the power switch flex cable through
the slot in the bezel.
2. Place the bezel onto the 6 bosses on the housing.