• To exit Advanced settings and return to the Home tab, touch Exit.
Specify advanced averaging interval settings
Averaging intervals are available in the Office profile.
1. In Advanced settings, touch the Parameters > Averaging tabs.
2. Specify settings.
Setting Action/Description
NIBP averaging program Select the averaging program to view or modify. You can
change the name of the averaging program.
Readings to average Select which readings in a series to use for averaging
(and the number of readings to average as a result).
Delay to start (minutes) Enter the number of minutes to wait before the
averaging program starts (after touching "Start
intervals") and the start of the first reading.
Time between (minutes) Enter the number of minutes to wait between readings.
Keep if + or – (mmHg) Enter the range that the program uses as criteria to
accept or reject readings and establish the baseline
reading. For more information about how this setting
affects the program, refer to the section "Excluded
readings" below.
Summary display box Displays currently selected settings.
3. Do one of the following:
• To continue in Advanced settings, touch another tab.
• To exit Advanced settings and return to the Home tab, touch Exit.
Specify advanced program interval settings
1. In Advanced settings, touch the Parameters > Program tabs.
2. Specify settings.
Setting Action/Description
Program Select the program to view or modify. You can change
the name of the program.
Intervals Specify up to five sets of interval readings for this
program. For each set, specify the number of minutes
between interval readings and the number of times that
each interval should run (repetitions) before proceeding
to the next set.
3. Do one of the following:
• To continue in Advanced settings, touch another tab.
• To exit Advanced settings and return to the Home tab, touch Exit.
Network tab
Specify radio settings
This task is applicable only to monitors that have a wireless radio installed. Devices can
be configured with a wireless or a Bluetooth radio.
26
Advanced settings Connex
®
Spot Monitor